Single-person O2CUBE Micro-pressure Oxygen Enrichment Chamber

Single-person O2CUBE Micro-pressure Oxygen Enrichment Chamber
+

Product Parameters

O2CUBE single-seat export model main parameters and technical description

Item

main parameters/ technical description

Cabin External Dimensions

2100mm*1050mm*1900mm(L*W*H,without wheels),transportable in sections, suitable for small home elevators

Cabin Internal Dimensions

2000mm*950mm*1800mm(L*W*H)

Floor Space

3.36m2

External Unit External Dimensions

≤650mm*600mm*1025mm(including wheels)

Power Supply

AC110V 50/60Hz

Maximum Power

2000W (including air conditioning)

Design Pressure

150kPa

Operating Pressure

1.1ata to 1.3ata, adjustable

Oxygen Concentration

≥90%

Oxygen Flow Rate

≥5L/min

Oxygen Inhalation Method

Headset-type oxygen inhalation device or nasal oxygen cannula oxygen inhalation device is optional

Oxygen Supply Mode

Dual oxygen supply modes, nasal inhalation and diffuse oxygen supply

the new national standard medical terminal

providing gentle oxygen delivery;

Pressure increase and reduction

The pressure increases and decreases at a gentle rate, providing a more comfortable experience

Real-time simulated altitude monitoring function

displays the simulated altitude in real time, adjustable from 0 to 5000 meters

Local atmospheric pressure adaptive function

automatically detects the local atmospheric pressure of the device

cabin

constructed of high-strength SPCC composite panels with a dedicated reinforcement structure, ensuring strength while reducing design complexity. The weight of the oxygen chamber is reduced, eliminating the load-bearing burden of traditional steel-structured oxygen chambers

The entire chamber

feature a modular design, allowing for easy access

Door

A self-locking mechanical door made of plexiglass, featuring lightweight, ≥92% transparency, and a light transmission dimension of ≥1600 x 600mm

Observation Window

A dedicated observation window with a pressure vessel-specific flange and high-performance, pressure-resistant plexiglass is used

Built-in independent air compressor unit

using a four-cylinder compressor with shorter piston stroke than a two-cylinder unit, resulting in low vibration, long service life, noise <50dB, high air output, and significant energy savings

Unique air treatment system

Incoming cabin air undergoes triple filtration and disinfection, features internal air circulation and odor removal, and automatic ventilation

The oxygen concentrator

utilizes imported high-efficiency lithium-based molecular sieves for stable performance

The molecular sieve adsorption tower in the oxygen concentrator

is constructed of one-piece aluminum alloy, with physical and chemical performance and technical indicators exceeding industry standards

intelligent lighting control system

allowing for a variety of atmosphere settings

An optional multimedia audio-visual system

is available in the cabin, enabling simultaneous wellness, healthcare, and entertainment

An advanced automation control system

equipped with a high-quality programmable logic controller (PLC), ensures fully automated and intelligent operation of the entire system. Energy-efficient operation

Equipped with dual color touchscreen human-machine interfaces

enabling one-touch operation and comprehensive information display

imported high-performance pressure transmitters

enabling real-time monitoring of cabin pressure

an optical oxygen concentration sensor

enabling real-time monitoring of cabin oxygen concentration, ensuring it remains within a safe and effective range

temperature and humidity detection devices

enabling real-time monitoring and control of cabin temperature and humidity

Ten safety protection measures

an automatic mechanical pressure relief valve, a dual-circuit automatic overpressure relief device, dual pressure detection systems, dual internal and external manual emergency pressure relief devices, high-strength, captive bolts for the cabin, a full-range equipotential bonding system, a cabin voltage limit of 24V, automatic short-circuit protection, automatic leakage protection, and a cabin oxygen concentration upper limit protection system

high-quality microfiber leather

flame-retardant and fire-resistant, ensuring safety and reliability

Multiple noise reduction technologies

addressing noise issues through both structural and material considerations, keeping cabin noise levels below 40 decibels

High-performance air conditioning units

safety and reliability

Triple purification

The first stage filters airborne dust particles larger than 5 μm; the second stage removes pollutants larger than 0.01 μm and removes odors; and the third stage uses molecular sieves to filter even finer particles

Environmentally friendly

The cabin's interior and exterior finishes are formaldehyde-free and volatile organic compounds-free, using materials that meet environmental standards

Our Advantages

Technical Proposal

Single-person O2CUBE hyperbaric oxygen chamber main configuration list

Serial number

name

Specifications and materials

unit

quantity

Remark

Cabin part

1

Micropressure oxygen chamber

2100mm*1050mm*1900mm

station

1

Sharded

2

air conditioner

commission

station

1

 

3

Functional sofa

Single seat

set

1

commission

4

Multimedia TV

32 inches

station

1

Optional

5

Oxygen terminal

Φ 8 single-pass

piece

1

 

6

Humidifier bottle

New national standard

piece

1

 

7

Nasal oxygen cannula

1.5 meters long

piece

2

 

8

silencer

commission

set

2

 

9

Wireless two-way intercom

commission

set

1

 

10

carpet

commission

set

1

 

11

Three-in-one module

commission

piece

1

 

12

touchscreen

commission

set

2

 

14

All-copper automatic air release valve

commission

set

1

 

15

interior and exterior decoration

commission

set

1

 

Outdoor unit

1

Microcompressed oxygen outboard machine

commission

set

1

 

2

Oxygen Generation System

≥5L/min

station

1

 

3

Four-cylinder air compressor

commission

set

1

 

4

PLC host

commission

set

1

 

5

PLC expansion modules

commission

set

1

 

6

PLC analog modules

commission

set

1

 

7

Electric ball valve

commission

set

1

 

8

solenoid valve

commission

set

1

 

9

Pressure transmitter

commission

set

1

 

Applicable Scenarios

Single-person O2CUBE Micro-pressure Oxygen Enrichment Chamber

Technical Proposal

Global Solutions